    Fluke 744 Calibrator Repair

    Time:2019-07-16             Xi'an Agitek spectrometer maintenance center    528
    1. Instrument model: FLUKE744 Documented Process Calibrator;
    2. Fault description: Press the power button and the machine cannot be turned on normally;
                             
    Fluke 744 Calibrator Repair
    3. Detection process:
    a) The output of the battery module is normal and the charging is normal;
    b) The key matrix group works normally;
    c) The internal power management module of the machine works abnormally, and the CPU&FLASH&CLOCK cannot be started normally;
    In summary, it is determined that the machine failure is caused by abnormal power supply.
    Four. Fluke 744 calibrator maintenance:
    Replace damaged devices and repair power management modules;
    Recalibrate the whole machine and adjust the indicators;
    The self-test & self-calibration passed, the machine returned to normal, and the maintenance was completed.